posted on 2023-02-22, 14:06authored byBarry Linnane, Donna Clarke, Colum P. DunneColum P. Dunne, Paula Murray, Niamh O'Sullivan, Paul McNally
Contamination of bronchoscopes is well
described.1–3 However, while cleaning
and disinfecting bronchoscopes are
clearly described in standardisation documents,
1–3 the performance of a control
sample prior to bronchoscopy is not suggested.
4 5 Here, we offer an argument
for such a procedure, supported by our
experience with a 2-year-old boy with
cystic fibrosis who underwent flexible
bronchoscopy and bronchoalveolar lavage
(BAL)