A validation study of a polymer optical fiber sensor for monitoring lumbar spine movement

Download (904.51 kB)
journal contribution
posted on 2022-10-03, 11:02 authored by WERN KAMWERN KAM, Mary O'Keeffe, KIERAN O'SULLIVANKIERAN O'SULLIVAN, Waleed S. Mohammed, Sinead O'KeeffeSinead O'Keeffe, Elfed LewisElfed Lewis, Charusluk Viphavakit
This study aims to investigate the validity and reliability of a novel plastic optical fiber (POF) sensor, which was developed to measure the angles of flexion, extension and lateral bend at the lumbar region. The angles of flexion, extension and lateral bend for a standing position were measured simultaneously using both the novel POF sensor of this investigation and the commercial Biometrics goniometer instrument. Each movement had two steps of bending which were 10 and 20 based on inclinometer readings. The POF sensor had good intra-rater reliability (Intraclass correlation coefficient, ICC = 0.61 to 0.83). Bland–Altman plots were used to study the agreement using these two sensors. There were proportional differences and bias between the POF sensor and Biometrics goniometer, as the zero points did not lie in the percentage difference region in the Bland–Altman plots. The proportional difference between these two likely reflects the different sizes and thus, measurement regions of the two sensors. There was also strong correlation between the two sensors (r > 0.77). Hence, the POF sensor could be of potential utility in measuring lumbar range of motion (ROM) in a manner which is minimally invasive, and where discrete sections of the spine are under specific investigation.

Funding

Study on Aerodynamic Characteristics Control of Slender Body Using Active Flow Control Technique

Japan Society for the Promotion of Science

Find out more...

History

Publication

Materials;12, 762

Publisher

MDPI

Note

peer-reviewed

Other Funding information

ERC

Language

English

Also affiliated with

  • Health Research Institute (HRI)

Department or School

  • Allied Health
  • Electronic & Computer Engineering

Usage metrics

    University of Limerick

    Categories

    No categories selected

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC